© 2026 Optics and Photonics at Nottingham
43%
56.3%
8.8%
£5m+
Data for 2020-2025 from SciVal
Swamy, Suvvi K Narayana; He, Chenyang; Hayes-Gill, Barrie R; Clark, Daniel J; Green, Sarah; Morgan, Stephen P
Pulse oximeter bench tests under different simulated skin tones Journal Article
In: Med. Biol. Eng. Comput., vol. 63, no. 7, pp. 1931–1942, 2025.
Abstract | Links | Altmetric | Tags: Melanin, Occult hypoxemia, oxygen saturation, Pulse oximeter, Racial bias, Skin colour, Transmission-mode
@article{Swamy2025-po,
title = {Pulse oximeter bench tests under different simulated skin tones},
author = {Suvvi K Narayana Swamy and Chenyang He and Barrie R Hayes-Gill and Daniel J Clark and Sarah Green and Stephen P Morgan},
doi = {10.1007/s11517-024-03091-2},
year = {2025},
date = {2025-07-01},
urldate = {2025-07-01},
journal = {Med. Biol. Eng. Comput.},
volume = {63},
number = {7},
pages = {1931\textendash1942},
publisher = {Springer Science and Business Media LLC},
abstract = {Pulse oximeters\' (POs) varying performance based on skin tones
has been highly publicised. Compared to arterial blood gas
analysis, POs tend to overestimate oxygen saturation (SpO2)
values for people with darker skin (occult hypoxemia). The
objective is to develop a test bench for assessing commercial
home and hospital-based POs in controlled laboratory conditions.
A laboratory simulator was used to mimic different SpO2 values
(~ 70 to 100%). Different neutral density and synthetic melanin
filters were used to reproduce low signal and varying melanin
attenuation levels. Six devices consisting of commercial home (Bioligh},
keywords = {Melanin, Occult hypoxemia, oxygen saturation, Pulse oximeter, Racial bias, Skin colour, Transmission-mode},
pubstate = {published},
tppubtype = {article}
}
Swamy, Suvvi K Narayana; Liu, Chong; Correia, Ricardo; Hayes-Gill, Barrie R; Morgan, Stephen P
Exploring the bias: how skin color influences oxygen saturation readings via Monte Carlo simulations Journal Article
In: J. Biomed. Opt., vol. 29, no. Suppl 3, pp. S33308, 2024.
Abstract | Links | Altmetric | Tags: Melanin, Monte Carlo, Occult hypoxemia, oxygen saturation, pulse oximeter, Racial bias, skin color, transmission mode
@article{Narayana_Swamy2024-la,
title = {Exploring the bias: how skin color influences oxygen saturation readings via Monte Carlo simulations},
author = {Suvvi K Narayana Swamy and Chong Liu and Ricardo Correia and Barrie R Hayes-Gill and Stephen P Morgan},
doi = {10.1117/1.JBO.29.S3.S33308},
year = {2024},
date = {2024-06-01},
urldate = {2024-06-01},
journal = {J. Biomed. Opt.},
volume = {29},
number = {Suppl 3},
pages = {S33308},
publisher = {SPIE-Intl Soc Optical Eng},
abstract = {Significance: Our goal is to understand the root cause of
reported oxygen saturation ( SpO 2 ) overestimation in heavily
pigmented skin types to devise solutions toward enabling equity
in pulse oximeter designs. Aim: We aim to gain theoretical
insights into the effect of skin tone on SpO 2 - R curves using
a three-dimensional, four-layer tissue model representing a
finger. Approach: A finger tissue model, comprising the
epidermis, dermis, two arteries, and a bone, was developed using
a Monte Carlo-based approach in the MCmatlab software. Two skin
tones-light and dark-were simulated by adjusting the absorption
and scattering properties within the epidermal layer. Following
this, SpO 2 - R curves were generated in various tissue
configurations, including transmission and reflection modes
using red and infrared wavelengths. In addition, the influence
of source-detector (SD) separation distances on both light and
dark skin tissue models was studied. Results: In transmission
mode, SpO 2 - R curves did not deviate with changes in skin
tones because both pulsatile and non-pulsatile terms experienced
equal attenuation at red and infrared wavelengths. However, in
reflection mode, measurable variations in SpO 2 - R curves were
evident. This was due to differential attenuation of the red
components, which resulted in a lower perfusion index at the red
wavelength in darker skin. As the SD separation increased, the
effect of skin tone on SpO 2 - R curves in reflection mode
became less pronounced, with the largest SD separation
exhibiting effects similar to those observed in transmission
mode. Conclusions: Monte Carlo simulations have demonstrated
that different light pathlengths within the tissue contribute to
the overestimation of SpO 2 in people with darker skin in
reflection mode pulse oximetry. Increasing the SD separation may
mitigate the effect of skin tone on SpO 2 readings. These trends
were not observed in transmission mode; however, further planned
research using more complex models of the tissue is essential.},
keywords = {Melanin, Monte Carlo, Occult hypoxemia, oxygen saturation, pulse oximeter, Racial bias, skin color, transmission mode},
pubstate = {published},
tppubtype = {article}
}
Ballaji, Hattan K; Correia, Ricardo; Korposh, Serhiy; Hayes-Gill, Barrie R; Hernandez, Francisco U; Salisbury, Byron; Morgan, Stephen P
A textile sleeve for monitoring oxygen saturation using multichannel optical fibre photoplethysmography Journal Article
In: Sensors (Basel), vol. 20, no. 22, pp. 6568, 2020.
Abstract | Tags: multichannel, oxygen saturation, photoplethysmography, plastic optical fibre, textile, wearable technology
@article{Ballaji2020-wp,
title = {A textile sleeve for monitoring oxygen saturation using
multichannel optical fibre photoplethysmography},
author = {Hattan K Ballaji and Ricardo Correia and Serhiy Korposh and Barrie R Hayes-Gill and Francisco U Hernandez and Byron Salisbury and Stephen P Morgan},
year = {2020},
date = {2020-11-01},
journal = {Sensors (Basel)},
volume = {20},
number = {22},
pages = {6568},
publisher = {MDPI AG},
abstract = {Textile-based systems are an attractive prospect for wearable
technology as they can provide monitoring of key physiological
parameters in a comfortable and unobtrusive form. A novel system
based on multichannel optical fibre sensor probes integrated
into a textile sleeve is described. The system measures the
photoplethysmogram (PPG) at two wavelengths (660 and 830 nm),
which is then used to calculate oxygen saturation (SpO2). In
order to achieve reliable measurement without adjusting the
position of the garment, four plastic optical fibre (POF) probes
are utilised to increase the likelihood that a high-quality PPG
is obtained due to at least one of the probes being positioned
over a blood vessel. Each probe transmits and receives light
into the skin to measure the PPG and SpO2. All POFs are
integrated in a stretchable textile sleeve with a circumference
of 15 cm to keep the sensor in contact with the subject's wrist
and to minimise motion artefacts. Tests on healthy volunteers
show that the multichannel PPG sensor faithfully provides an
SpO2 reading in at least one of the four sensor channels in all
cases with no need for adjusting the position of the sleeve.
This could not be achieved using a single sensor alone. The
multichannel sensor is used to monitor the SpO2 of 10
participants with an average wrist circumference of 16.0 $±$
0.6 cm. Comparing the developed sensor's SpO2 readings to a
reference commercial oximeter (reflectance Masimo Radical-7)
illustrates that the mean difference between the two sensors'
readings is -0.03%, the upper limit of agreement (LOA) is
0.52% and the lower LOA is -0.58%. This multichannel sensor
has the potential to achieve reliable, unobtrusive and
comfortable textile-based monitoring of both heart rate and SpO2
during everyday life.},
keywords = {multichannel, oxygen saturation, photoplethysmography, plastic optical fibre, textile, wearable technology},
pubstate = {published},
tppubtype = {article}
}
A part of the University of Nottingham