Ma, Zi-Yu; Sun, Shen; Wu, Shui-Cai; Lin, Lan; Chen, Yi-Xiong; Zhao, Dong; Morgan, Stephen P
Pilot study on optimizing pressure for standardized capillary refill time measurement Journal Article
In: Heliyon, vol. 10, no. 15, pp. e35716, 2024.
Abstract | Links | Altmetric | Tags: Capillary refill time, Microcirculation, Pressure, Sepsis, Septic shock
@article{Ma2024-lq,
title = {Pilot study on optimizing pressure for standardized capillary refill time measurement},
author = {Zi-Yu Ma and Shen Sun and Shui-Cai Wu and Lan Lin and Yi-Xiong Chen and Dong Zhao and Stephen P Morgan},
doi = {10.1016/j.heliyon.2024.e35716},
year = {2024},
date = {2024-08-01},
urldate = {2024-08-01},
journal = {Heliyon},
volume = {10},
number = {15},
pages = {e35716},
publisher = {Elsevier BV},
abstract = {Purpose: Capillary Refill Time (CRT) measurement has gained
increasing attention in the field of sepsis and septic shock.
Recognizing pressure as a fundamental determinant in CRT
measurement is crucial for establishing a standardized CRT
measurement procedure. In this preliminary study, we elucidated
the optimal pressing strength for CRT measurement by analyzing
the CRTs measured under varying pressures. Method: Seventeen
healthy individuals were enlisted to undergo CRT tests on their
fingertips at various pressure levels. The applied force was
initiated at 0.5N and incrementally increased by 0.5N until it
reached 10.5N. An integrated Photoplethysmography (PPG) device
was employed to capture fluctuations in light intensity. The CRT
was automatically derived from the PPG signals via a specialized
algorithm. The study included correlation assessment and
reliability evaluation. Box plot and Bland-Altman plot were used
to visualize the impact of pressure levels on CRTs. Results: A
dataset of 1414 CRTs across 21 pressures showed significant
differences (Kruskal-Wallis test, p \< 0.0001), highlighting the
impact of pressure on CRT. CRT values between 4.5N and 10.5N
pressures varied less, with an Intraclass Correlation
Coefficient (ICC) of 0.499 indicating moderate consistency.
Notably, CRTs at 10N and 10.5N pressures revealed a high ICC of
0.790, suggesting strong agreement. Conclusion: A pressure range
of 4.5N-10.5N is recommended for stable CRT measurements, with
10.0N-10.5N providing optimal consistency and reliability.},
keywords = {Capillary refill time, Microcirculation, Pressure, Sepsis, Septic shock},
pubstate = {published},
tppubtype = {article}
}